BASILICA (leaflet laceration before TAVI): a Cath Lab 101 guide
Reviewed by Rodnie Oro, cath lab charge nurse · Last reviewed: October 2026 · Education only, not medical advice
BASILICA in one minute
What it is: BASILICA (Bioprosthetic or native Aortic Scallop Intentional Laceration to prevent Iatrogenic Coronary Artery obstruction) is a catheter technique done just before TAVI. An electrified wire splits (lacerates) an aortic valve leaflet down the middle, so that when the new valve pushes the old leaflets aside, the split leaflet opens like a curtain instead of covering the coronary artery opening.
Who it is for: people having TAVI, especially valve-in-valve TAVI inside a failed surgical valve, whose CT shows a high risk that a leaflet could block a coronary artery, as decided by a heart team. It is a specialist technique done in experienced centres.
Main risks: stroke, damage to the aortic root or valve, coronary obstruction despite the procedure, severe leak before the new valve is placed, and the usual TAVI risks. Your team will explain your own risk.
Usually: under general anaesthetic or sedation with TAVI in the same procedure, about 2 to 3 hours in total, often home after a few nights.
On this page: What is BASILICA? · Who is it for? · When is it done? · Step by step · Recovery · Risks · Kit list · First time in the lab? · FAQ
Coronary obstruction risk, cut-away of the aortic root: (A) the old leaflet, pushed aside by the new valve, can cover the coronary opening; (B) after BASILICA the split leaflet no longer covers it.
What is BASILICA?
During TAVI, the new valve is expanded inside the old one and pushes the old leaflets outwards against the aortic wall. If a leaflet is long, bulky or sits close to a coronary artery opening (especially inside a surgical valve, or where the aortic root is narrow), it can cover the coronary opening, cutting off blood flow to the heart muscle. This coronary obstruction is rare but life-threatening.
BASILICA prevents this by splitting the leaflet in front of the at-risk coronary artery before the new valve goes in:
a wire is passed through the base of the leaflet using a brief burst of electrosurgical energy
the wire is caught on the other side and folded into a V shape
pulling the energised wire splits the leaflet from base to tip
when the TAVI valve expands, the two halves of the split leaflet splay apart, leaving a channel for blood to reach the coronary artery
BASILICA can be done on one or both leaflets in front of the coronary arteries.
New to the acronyms? See the Cath Lab Abbreviations A–Z.
Seen from above (aorta towards the heart): (1) the wire crosses the base of the leaflet in front of the at-risk coronary; (2) after BASILICA and TAVI the leaflet is split and splays apart, so the coronary is uncovered.
After BASILICA and TAVI the split leaflet splays apart against the wall: (1) one half; (2) the other half; (3) the gap between them, so blood can still reach the coronary artery.
Who is it for?
The ESC/EACTS valvular heart disease guidelines and NICE guidance on valve-in-valve TAVI (HTG518) recognise coronary obstruction as a risk to be assessed by CT before TAVI. Techniques such as BASILICA, or alternatives such as coronary protection with a stent, are considered by an experienced heart team when the CT shows a high risk. BASILICA is a specialist, evolving technique, and consent includes discussion of this.
Risk is assessed by CT, measuring how close the leaflet will sit to the coronary opening after TAVI and the width of the aortic root.
When is it done?
It is planned with the TAVI. Before it, the team usually arranges:
detailed CT planning of the aortic root, leaflets and coronary arteries
echo, ECG, blood tests and a group and save
a review of blood thinners, as per centre protocol
Step by step
1. Anaesthetic and access
Done under general anaesthetic or sedation, often with TOE. Sheaths are placed in both femoral arteries (one for the laceration catheters, one for the snare and later the TAVI sheath) and a femoral vein for temporary pacing. Sizes per device and operator. Heparin is given, dose as per centre protocol.
Femoral access: (1) catheter for the laceration wire, right femoral artery; (2) catheter holding the snare, left femoral artery; (3) pacing wire in the femoral vein; (4) the aortic root and the target leaflet.
2. Leaflet laceration
A. Set up. A snare is placed in the left ventricular outflow tract below the target leaflet. From the other side, a guide catheter with an inner catheter is aimed at the base of the leaflet.
B. Traverse. A thin wire, insulated except at its tip and connected to an electrosurgery pen, crosses the leaflet with a brief burst of energy (settings per operator) while non-conducting fluid (such as 5% dextrose, as per centre protocol) is flushed through the guide. The wire is caught by the snare.
C. Lacerate. The wire is pulled out through the other sheath, its middle is bent into a V (the "flying V") and positioned at the leaflet base. Both ends are pulled while the wire is energised, splitting the leaflet from base to tip.
D. TAVI. The TAVI valve is implanted straight away in the usual way, because the split leaflet causes a leak until then.
A: the wire crosses the base of the leaflet and is snared. B: the wire is pulled through with energy on, splitting the leaflet from base to free edge.
A: guide catheter at the leaflet and a snare waiting in the LVOT. B: the wire passes through the base of the leaflet and is snared. C: the energised wire is pulled through the leaflet (the flying V). D: TAVI; the split leaflet splays apart and the coronary stays open.
3. Checking
An aortogram, echo and ECG check that both coronary arteries fill and the new valve works well.
Illustrated CT planning, not a patient scan. The dashed box is the planned valve; the arrow is the predicted distance between the pushed-aside leaflet and the coronary opening. A short distance, for example under about 4 mm, suggests a higher risk. Thresholds are examples.
Recovery and aftercare
First hours: recovery, often in coronary care or a high-dependency area, with checks of the groins, heart rhythm (including for heart block) and blood pressure, as per centre protocol.
Echo and blood tests: before discharge.
Going home: often after a few nights.
Medicines: antiplatelet or anticoagulant medicines, as per your team's plan.
Driving: DVLA guidance after a percutaneous valve procedure for a car licence is not to drive for 2 weeks; bus and lorry drivers must tell the DVLA. Ask your team.
Call 999 if you have chest pain, sudden breathlessness, fainting, signs of a stroke, or bleeding from the groin.
After BASILICA and TAVI: (1) monitoring; (2) echo and bloods; (3) home after a few nights; (4) follow-up at 6 to 10 weeks.
Risks and complications
Your team will discuss your own risks. Possible problems include:
stroke, from debris or air
coronary obstruction despite the laceration
injury to the aortic root, the mitral valve or other structures from the wire
a severe leak between laceration and TAVI, which is why the valve follows immediately
bleeding or injury at the access arteries
the usual TAVI risks, including heart block needing a pacemaker
Cath lab kit list (printable)
Internal mammary and multipurpose diagnostic catheters; pigtail catheters (from the guide kit list; no published kit-page picture)
Barts example kit list
An example from one UK centre, written generically. Every lab has its own protocol, and sizes and doses vary.
Access
femoral arterial sheaths on both sides, and a femoral vein sheath for pacing; large-bore TAVI sheath; sizes per device and operator
Catheters
guide catheters (extra-backup, right coronary and left coronary/AL shapes, one on standby), per operator
internal mammary and multipurpose diagnostic catheters; pigtail catheters
microcatheter or wire converter, per operator
Wires and snare
stiff-tipped 0.014" traversal wire (to be insulated except the tip), per operator
0.014" buddy wire; extra-stiff wires; stiff pre-shaped LV wire, per operator
snare (20–25 mm class, per operator) with haemostasis valve
Electrosurgery
diathermy pen and generator, settings per operator
5% dextrose for flushing during energy delivery, as per centre protocol; one-way valve or needle-free connector
haemostasis valves
Usual TAVI equipment
Procedural steps (Barts example)
Both femoral arteries and a femoral vein; pacing wire; heparin
Snare (via a right coronary-shape catheter) in the LVOT under the target leaflet
From the other side: guide catheter, inner catheter and traversal wire aimed at the leaflet base
Wire connected to the diathermy pen; dextrose flush; energy applied while advancing; wire snared
Wire externalised; flying-V formed and positioned; energised pull splits the leaflet
TAVI performed immediately; coronary flow and valve function checked
First time in the cath lab?
đź’ˇ First-time tip
Dextrose, not saline. During energy delivery the guide is flushed with dextrose, as per centre protocol. Label the syringes clearly. Diathermy only on command. The pen is activated only when the operator says so. Have the TAVI ready. The valve must be prepared before the laceration, because it goes in straight away. Watch the pressure. A big leak after the laceration can drop the blood pressure; the TAVI fixes it. Two snares, many catheters. Keep the table organised; label which catheter is which side.
FAQ
Why not just do the TAVI?
If CT shows a high risk that a leaflet could block a coronary artery, BASILICA reduces that risk.
Does it make TAVI longer?
Yes, a little. It is done in the same procedure.
Is it widely done?
It is offered in experienced centres for selected patients.
Are there alternatives?
Yes, such as placing a stent in the coronary artery during TAVI, or surgery. Your team will discuss them.
Related Cath Lab 101 guides
Sources
All checked on 10 October 2026. Each page's own date is shown where it gives one.
NICE. Valve-in-valve TAVI for aortic bioprosthetic valve dysfunction (HTG518). nice.org.uk/guidance/htg518 (published 26 Jun 2019)
NICE. Transcatheter aortic valve implantation for aortic stenosis (HTG446). nice.org.uk/guidance/htg446
Vahanian A, et al. 2021 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J 2022;43(7):561–632. PubMed 34453165
British Heart Foundation. TAVI. bhf.org.uk
DVLA / GOV.UK. Cardiovascular disorders: assessing fitness to drive. gov.uk
Disclaimer
Education only. This page is for general education for healthcare students, cath lab staff and curious readers. It isn't medical advice and doesn't replace advice from your own doctor, heart team or local unit protocols. Equipment and medicines are described generically. No product or manufacturer is recommended. If you think someone is having a heart emergency, call 999.
Disclosure: I work for a company that makes a device used in this procedure. This page is independent education, isn't sponsored or reviewed by any company, and doesn't recommend any product.
Diagrams are original illustrations created for rodnieoro.com, except procedure images credited in their captions.